Kurfürstenstr. 75-79
13467 Berlin, Germany
Phone: +49 30 4044087
Fax: +49 30 4040529
Manufacturers
Company Figures
Number of employees 1-19
Sales volume 1-9 Mio US $
Export content > 75%
Year of foundation 1946
Area of business ■Laboratory Equipment
Company Profile
About us
COMPANY NAME
ROBERT RIELE GmbH & Co KG
MANAGING DIRECTOR
Mr Dipl.-Ing. Wolfgang Riele
(university degree in Electrical Engineering)
W.Riele@riele.de
EXPORT MANAGER
OFFICER WITH STATUTORY AUTHORITY
Mrs Dipl.-Kauffrau Isabel Riele
(university degree in Business Administration)
Isabel.Riele@riele.de
ADDRESS
Kurfuerstenstrasse 75-79
13467 BERLIN
GERMANY
TELEPHONE
TELEFAX
E-MAIL
+49 (0)30 404 40 87
+49 (0)30 404 05 29
info@riele.de
INTERNET
www.riele.de
REGISTRATION NO
AG Charlottenburg, HRA 1773
FULLY LIABLE PARTNER
Riele Beteiligungs- und Verwaltungs-GmbH
REGISTRATION NO
AG Charlottenburg, HRB 7143
YEAR OF FOUNDATION
1946
MEMBERSHIPS
Industrie- und Handelskammer zu Berlin
(Chamber of Commerce)
Allgemeiner Verband der Wirtschaft für Berlin und Brandenburg e.V.
(employers’ federation)
Berufsgenossenschaft der Feinmechanik und Elektrotechnik
(professional association)
BANK
Deutsche Bank
10585 Berlin
Telex: 1 81 433 dba
BIC (SWIFT-Code): DEUT DE BB
account no. 531 32 00 00
IBAN: DE63 1007 0000 0531 3200 00
VAT NO
DE 136 65 53 11
CERTIFICATES
DIN EN ISO 9001 (since July 1997)
DIN EN ISO 13485 (since Sep 2003)
ACTIVITIES
Design, production, sales and servicing of analysing devices, especially photometers, OEM
REFERENCES
Minitüb
eagle lab’s
Boehringer Mannheim, Behring, Merck, WTW, Macherey-Nagel
Product Information Product Categories Product
Information
Analyser appliances
Filter photometers
Cuvettes for glass photometers
Cuvettes for synthetic photometers
Photometers 3 Products
Accessories for photometers
Temperature regulation equipment
Dry warm up devices
Bilirubinometers/bilimeters
Devices and systems for glucose determination units
Glucose photometric determination
Devices and systems for haemoglobin determination
Units and systems for clinical chemistry
Semi-automatic units for clinical chemistry
Mechanical units for clinical chemistry
Product Information of Product Category Photometers Photometer 4040 V5+ 21.08.2009 Type
Single-beam filter photometer
Light Source
Halogen lamp – 12 V, 20 W with protection for lifetime
Wavelength
340 nm – 800 nm
Wavelength Selection
Automatic via 9-position filter wheel :
6 standard interference filters: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm, and 623 nm;
3 positions for optional filters of choice
Photometric Range
0 – 2.5 A
Cuvette System
Normal standard cuvettes (macro or semi-micro, disposable or special optical glass)
Temperature Control
Measuring cell as well as dry incubation block with 8 positions temperature controlled by electrical thermostat at 37 °C
Operator Interface
Touchscreen for direct functions and alphanumerical inputs
Data Presentation
Graphic display: White characters or symbols, blue background, lighted, resolution 240 * 128 dots
Integrated Printer
Dot-matrix printer, 24 characters per line
Languages
English and German / Indonesian / Russian / Spanish
Memory
•General operating software can be updated by PC
•Reagent-open system with capacity for up to 231 programmable methods
•Data import by touchscreen or PC
•Up to 50 nonlinear calibration curves with max 20 sets of points can be stored
Patient Data Management
Up to 2970 results of max 99 patients can be managed in the memory
Signal Port
Serial port for connection to an external printer and/or PC
Data Logging
Up to 2970 results of max 99 samples can be managed in the memory
Measurement
Procedures
•Absorbance, Transmission
•Endpoints with factor, standard or multiple standards, with or without reagent blank and/or sample blank
•Bichromatic end point
•Kinetics with standard, factor or multiple standards, with or without blank
•Fixed time with standard, factor or multiple standards, with or without reagent blank
•Turbidimetry with optional timer function
•Single, double and triple determinations
•Curve fitting for nonlinear standard curves
•Free hemoglobin in combination with optional interference filters
Quality Control
Up to 50 methods can be controlled with two control serums, Levey Jennings plot
Measuring Time
•Kinetic: variable from 3 – 25 deltas, time per delta 4 – 255 s
•Fixed time: variable from 0 – 65535 s
Delay Time
Programmable from 0 – 65535 s
Mains Supply
Range: 90 VAC up to 264 VAC at 50/60 Hz
Dimensions
Length 33 cm x width 34 cm x height 18 cm
Weight
6 kg
Photometer 5010 V5+ 21.08.2009 Type
Semi-automatic, single-beam filter photometer
Light Source
Halogen lamp – 12 V, 20 W with protection for lifetime
Wavelength
340 nm – 800 nm
Wavelength Selection
Automatic via 9-position filter wheel :
6 standard interference filters: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm and 623 nm;
3 positions for optional filter of choice
Photometric Range
0 – 2.5 A
Cuvette System
Micro flow cell: 32 µl, 10 mm light path interchangeable with normal standard cuvettes (macro or semi-micro, disposable or special optical glass)
Temperature Control
•Internal Peltier element, temperature variable, pre-adjusted to 25 °C, 30 °C and 37 °C
•Equilibration time for aspirated reaction mixture to reach 37 °C from ambient temperature: 15 sec
Aspiration System
Built-in peristaltic pump driven by stepper motor, programmable aspiration volume controlled by infrared light barrier
Sipping Volume
•Minimum 250 µl, typically 500 µl up to 2000 µl
•Separate setting of aspirate volume and wash volume
Operator Interface
Touchscreen for direct functions and alphanumerical inputs
Data Presentation
Graphic display: white characters or symbols, blue backlight, dimension 240 * 128 dots
Integrated Printer
Dot-matrix printer, 24 characters per line
Languages
English and German / Indonesian / Russian / Spanish
Memory
•General operating software can be updated by PC
•Reagent open system with capacity for up to 231 programmable methods
•Import of data by touchscreen or PC
•Up to 50 non-linear calibration curves with max 20 sets of points can be stored
Data Logging
Up to 2970 results of max 99 samples can be managed in the memory
Signal Port
Serial port for connection to an external printer and/or PC
Measurement Procedures
•Absorbance
•End Point with factor, standard or multiple standards, with or without reagent blank and/or sample blank
•Bichromatic end point
•Kinetics with standard, factor or multiple standards, with or without blank
•Fixed time with standard, factor or multiple standards, with or without reagent blank
•Turbidimetry
•Single, double and triple determinations
•Curve fitting for non-linear standard curves
•Free hemoglobin in combination with optional interference filters
Quality control
Up to 50 methods can be controlled with two control sera
Levey Jennings plot
Measuring Time
•Kinetic: variable from 3 – 25 deltas, time per delta 4 – 255 sec in 1 sec steps
•Fixed time: variable from 0 – 65535 sec in 1 sec steps
Delay Time
Programmable from 0 – 65535 sec in 1 sec steps
Mains Supply
Range: 90 VAC up to 264 VAC at 50/60 Hz
Dimensions
Length 33 cm x width 34 cm x height 18 cm
Weight
6 kg
Photometer 660 21.08.2009 made in Germany for analyses of substrates, electrolytes, lipids and proteins in the small lab.
•no lamp change (LED)
•no battery change (rechargeable)
•mobile by dual power supply (battery/mains)
•easy to use by pre-programmed parameters
•open system
Type
Single-beam filter photometer
Light Source
LED, extremely bright white
Cuvette System
Square cuvette 10 mm or 16 mm Ø tubes
Optical filters
492, 546, 578 nm, 4 positions for optional filters, manually driven filter wheel, position controlled
Display
LCD, 2 lines of 16 characters
Keyboard
16 keys
Measurement procedures
Absorbance, endpoint with factor or standard, transmission, curve fitting for non-linear standard curves
Interface port
Serial port for printer, PC, EDP
Power
Battery rechargeable: NiCd 7.2 V / 1600 mAh
Mains: charger 230 VAC
Dimensions
W 240 x D 215 x H 82 mm
Weight
Instrument 1.5 kg
Charger 230 g
Contact Persons Department/ Name Contact / e-Mail
Vertrieb/Isabel Riele Isabel.Riele@riele.de